Main Feature
l Adopt high-resolution thermal array output system(8 dot/mm), no need for any
adjustment,recording frequency: up to 150Hz
l Real-time and continuously record the clear and accutare single-channel ECG waveform and
annotation character(including lear mark, sensitivity, paper speed and filter status, etc)
l In auto print mode, finish recording with a button, which improves the work efficiency.
l Control all operations with buttons, convenient in operating, display work status with LCD,clear and
legible in equipmen state.
l AC and DC are optional, in optional DC state, built-in rechargeable lithium battery can contiuously
working for 10 hours, print up to 3 hours and 160-ECG, which meets the requirement for visiting a
patient at home and examination
l Adopt digital single processor, and get higher quality ECG waveform via AC filter, baseline filter and
EMG filter of ECG single.
Suitable For
Technical Specifications
Safety Classification: Class I, type CF
Input mode: Floating, Defibrillation protection and pacemaker pulse rejection circuit
Frequency response: 0.05Hz~150Hz (-3dB~+0.4dB)
CMRR: >60dB, >100dB, (add filter)
Time constant:≥3.2s
Patient leak current: < 10uA
Standard sensitivity:10mm/mV±2%
Noise level:≤15uVp-p
Input Impedance:≥50MΩ
Recording mode: thermal printing system
Specification for recording paper:50mm(W)*20m(L),high-speed thermal paper
Polarization voltage:±300mV
EMG interference filter: 35Hz(-3dB)
Paper speed: 25mm/s, 50mm/s, ±5%
Fuse specification: AC time-delay fuses, (5*20mm)T1.6AL250V
Dimension:315mm*215mm*77mm
Power supply: AC 100~240v, 50/60Hz ; DC 7.4V,2000mAh lithium battery
Accessories
Lead cable
Veterinary ECG clip
Thermal recording paper
Power cord
Earth wire
User manual